Headlines News :
Welcome To Electrical Laboratory PT.Qualis Indonesia

Oxygen & Air Bomb Aging Chamber

General Information
The Oxygen & Air Bomb Aging Chamber is designed according to the standard IEC 60811-1-2:1985 Amendment 2 2000-07 (Common test methods for insulating and sheathing materials of electric and optical part 1-2: Methods).
The standard of IEC 60811-1-2:1985 specifies the test methods to be used for testing polymeric insulating and sheathing materials of electric cables for power distribution and telecommunications including cables used on ships. It supplies the thermal aging methods which apply to the most common types of insulating and sheathing compounds (Elastomeric, PVC, PE, PP, and ect.)
Kind of test using the equipment:
- Ageing in an air bomb test. (IEC 60811-1-2, SNI 04-6190.1.2)  
- Ageing in an oxygen bomb test. (IEC 60811-1-2, SNI 04-6190.1.2)  

Features  
- Release of pressure (manually or auto).  
- Release of pressure speed (fast or slow)  

Technical Specifications  
- The maximum work pressure: 2.5 MPa  
- Safe blow up pressure: 2.7 ~ 3 MPa  
- Pressure meter:  
  - Range: 0 ~ 4 MPa  
  - Resolution: 0.02 MPa  
- Temperature controller:  
  - Range: RT ~ 150  
  - Resolution: 0.1  
- Fluctuation: 0.5  
- Effective volume: ≥4000 cm3 
- Air bomb:  
  - Test temperature: 127  or (ruled according other standards).  
  - Test pressure: 0.55 ± 0.02 MPa  
- Oxygen bomb:  
  - Oxygen not less than 97% purity.  
  - Test temperature: 70 (or ruled according other standards).  
  - Test pressure: 2.1 ± 0.07 MPa
- Timer: 0.015 ~ 9999 H

Copyright 2013 PT.Qualis Indonesia